Xantrex DC XG 850 User Manual Page 39

  • Download
  • Add to my manuals
  • Print
  • Page
    / 286
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 38
Step 5: Performing Functional Tests
M370078-01 2-9
2
Voltage and Current Mode Operation Checks
To perform the voltage and current mode operation checks:
1. Ensure that the front panel power switch is in the On position and the
output is disconnected.
2. If the OUTPUT ENABLE Main button is illuminated, press the
button to turn off the output.
3. To check voltage mode operation, turn the 9-position mode control to
the VOLTS position.
The voltage set point will blink dimming and then return to full
brightness. For more information, see
“Local Operation” on page 3–1.
4. Adjust the voltage to 5 V.
5. To check current mode operation, turn the 9-position mode control to
AMPS position.
Verify that the current set point is blinking in the output current
display.
6. Adjust the current to 1 A.
7. Press the OUTPUT ENABLE Main button to turn On.
8. Turn the front panel power switch to the Off position.
9. Turn the front panel power switch to the On position.
10. Connect a short circuit across the output terminals. Use leads of
sufficient current carrying capacity. (See
“Step 4: Selecting Load
Wires.)
11. Press the OUTPUT ENABLE Main button to enable the output. The
button will be illuminated when the output is enabled.
CC Mode LED illuminates and the voltage and current are displayed.
CV Mode LED illuminates and the preset load current is displayed.
12. Turn the front panel power switch to the Off position.
Page view 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 285 286

Comments to this Manuals

No comments