Alarms and Errors
M370078-01 3-21
3
Alarm Masking
It is possible to completely disable some alarms through the use of the
alarm mask. If an alarm is masked then this masking will prevent it from
registering in the SCPI conditions registers as well as not triggering the
alarm. The alarms that can be masked are identified in
Table 3-4 on
page 3–18. Setting the bit position (1) of an alarm will result in the alarm
being enabled. Clearing the bit position (0) of an alarm will result in the
alarm being masked/disabled.
For Example:
If you wanted to enable the OTP, Foldback and UVP alarms you
would take the OTP value of 2 and add it to the foldback value of 256
and the UVP value of 1024 to get a mask of 1282. This is the value
you would send with the SCPI command.
Control of the alarm mask is only available using the SCPI command.
The SCPI command (s) for these instructions are:
[:]SYSTem[<channel>]:PROTection:MASK
The mask command takes a single parameter made up of the sum of the
values for the alarms that are to be masked. See
Table 3-5 for the alarm
values associated with each maskable alarm.
The SCPI command (s) for these instructions are:
[:]SYSTem[<channel>]:PROTection:MASK
Table 3-5
Alarm Mask Bit Positions
Alarm Bit Position Value
Over Temperature Protection (OTP) 1 2
Fan 3 8
Foldback 8 256
External Shutdown 9 512
Under Voltage Protection (UVP) 10 1024